Medical Device
Translation And Compliance
Any change to the design
should be treated in the same way as the creation of
the design. In such a case, all the necessary steps
need to be followed, including reviews, verification,
validation, approvals, and documentation. Once a
product is released, any new change in the product
design can affect all the users and badly reflect on
the reputation of the manufacturer. So, the design
changes are to be made at the earliest step and
implemented before release.
Testing Strategies
In order to get market
acceptability, every medical device must meet the
functionality, usability, and reliability objectives.
Apart from these, each device addresses a particular
problem or condition, some of which are critical to
life. Therefore, it is important to prove the
effectiveness and safety of devices to the end users.
This is achieved through iterative testing during
verification and validation processes. These are
driven by the regulatory environment and must follow
international standards. The set of International
Standards under ISO 10993 - "Biological evaluation of
medical devices"- provides a reliable and globally
accepted roadmap for the various tests and analyses
related to medical devices.
The test strategy should
be planned considering – (i) The device segment and
product class (ii) Prescribed standards for the
product as well as for testing, (iii) Capability of
the testing labs and their accreditation status, (iv)
The sequence of tests and the duration of each, (v)
Total cost and the turn-out time of testing, and (vi)
Final documentation for the regulatory requirement.
Being an iterative
process, the testing is money-intensive and
time-consuming. A strongly-defined testing/evaluation
strategy helps to optimize cost as well as the test
period to make the product market-ready on time. Also,
the testing activity helps to comply with regulations
as well as designing the highest quality product and
manufacturing processes.
Verification and
validation processes executed through lab tests, help
to ensure that the device is aligned with the needs of
targeted users and that it delivers the intended
solution. Also, they ensure that all the requirements
are met or not, and enhance the approval process.
Risk Management
Procedures
Risk Management is the
process of identifying, controlling, and preventing
the failure that may cause hazards to users. It also
mandates identifying associated risks. Upon reaching
an unacceptable level of risk, it notifies developers
to decrease at least up to an acceptable level.
Basically, the risk
management of medical devices is enforced by the
international compliance standard ISO 14971:2007
Medical Devices – “Application of Risk Management to
Medical Devices”. Apart from this, risk management
policies need to be incorporated across all the stages
of medical device design and development. Evidently,
it should be integrally linked with the design control
process. Given the complexity of developing a medical
device, risk management practices help ensure
usability, safety, and regulatory compliance.
Figure 2 : The risk
management process.
Figure 2 shows the
modalities involved in the risk management process.
The process starts with the identification of all
hazards. Then select a specific one and measure the
associated risks based on the consequences of hazards
and the probability of risk. If the identified risk
level is above the defined criteria, then it needs to
be mitigated. The risk level depends on many
parameters such as the device, technologies, or even
the risk acceptability policy. Risk mitigation
strategies could be implemented till the risk level
becomes acceptable. Once the specific hazard becomes
acceptable in terms of the risks, then the next hazard
could be analysed.
Before finalizing a
design in the medical device development, it is good
practice to conduct a hazard analysis to get an idea
about the standard hazards associated with the device.
The ‘Primary Hazard Analysis’ can be done considering
the following hazards :
(I) Raw materials and
wastes: Toxicity, flammability, and reactivity of
material.
(ii) Environmental
factors: Sensitivity to temperature, humidity etc. Any
environment pollution during the process.
(iii) Mechanical or
electronic hazards : Whether the device (or its
components) is likely to pose any health threat to the
user, like electric shock or injury.
(iv) User device
interface: Hazards associated with human factors like
ineffective delivery, drug administration, incorrect
or incomplete information, control of life-sustaining
operations
When multiple hazards are
identified, they can be prioritized according to the
severity associated with them.
Practical Aspects of
Medical Device Development
The Design Control
Process described in the earlier section is mainly
related to the compliance aspects of medical device
development. Apart from that, a manufacturer may have
to go through 5 distinct phases related to the product
lifetime.
Phase 1: Device discovery
and risk analysis
Phase 2: Formulation, concept, and feasibility
Phase 3: Design and development - verification and
validation
Phase 4: Final validation and product launch
preparation
Phase 5: Production, market introduction, and
post-market follow-up
|