Calibration of Electronic
balances and mass
Class E2 reference mass sets are
used in the calibration of electronic balances and
mass sets. OIML guidelines and ASTM standards are
followed in the execution of calibration procedures.
Pipettes and
volumetric glass wares
Volume calibrations are
performed using electronic balances. Volumetric
apparatus ranging from micropipettes to volumetric
flask and burettes are being calibrated at CAC.
Calibration of
Rotational speed
Non contact stroboscopes
are available as reference standard for the
calibration of centrifuges.
Thermometers,
Incubators-calibrations
ITS-90 calibrated
Secondary Platinum Resistance Thermometers are used as
the reference equipment in temperature calibrations.
Comparison technique is the method followed in the
calibration.
Relative Humidity (RH)
Calibrations
Controlled environment
such as temperature and humidity conditions are
significant in performing tests and evaluations, as
they may affect results. Calibrated RH meters are used
for routine monitoring of environmental conditions.
Saturated salt bath solutions and calibrated RH meters
at controlled environment makes the system for RH
meter calibrations.
Electro technical
calibrations, calibration of UV-Visible
spectrophotometers and Sterilizers are some of the
other important activities at CAC.
Facility for
Characterization of Surface finish
Surface finish of
biomaterials have an important role in their
performance when they form a part of medical device. A
noncontact and non destructive surface profilometer
helps in characterizing without damaging precious
samples. CLI gauges based on principles of chromatic
length aberration and laser gauges are available for
the non - contact measurements.
|