A TECHNO-ECONOMIC NEWS MAGAZINE FOR MEDICAL PLASTICS AND PHARMACEUTICAL INDUSTRY
Our 10th Year of Publication
Page 02

Qulaity

Evaluation of Medical Devices

Dr. K.V.Jogi Director, DCTL Mumbai

Testing is done mainly to verify the claims or to validate the system to meet product specifications.

Test results are functions of operating environment and procedure adopted and the data generated should be interpreted in these contexts.

Tests may be categorized as:

  1. Time testing to determine long-time reliability by estimating failure rate and mean time between failures and can show which part or component cause failure and also, mechanism of failure.

  2. Event Testing is repeated testing, till failure.

  3. Stress Test done by applying limiting stress that may be encountered in normal usage as well as possible (usual) mishandling. Cause of failure may be investigated by step by step evaluation.

  4. Environmental Testing is done by exposing the device to environmental and transport conditions (temperature, humidity, vibration, electromagnetic radiation etc.), that are normally encountered.

Time related: Test conditions are applied for certain period of time.

Failure related: Tests conducted till one or more devices fail.

Purpose of Test determines the choice and methods of test.

Design verification requires most extensive testing of every minute stage or component to verify the output conforms to the input and the parameters selected would be determined by the expected output.

Tests during manufacture are so selected as to ensure consistency of the component and the system to predetermined specifications likely to be crossed due to random variations.

Tests for acceptance are usually limited to parameters most likely to affect functioning of the device.

Investigation of failure in the field will be determined by the type of failure.

Another important area of Evaluation is Biocompatibility. Devices which directly or indirectly came into contact with body of the consumer would have to be tested. These are designed to assess potential adverse effects in conditions of actual use or close to them.

Types of Tests and parameters of study depend on the tissue and duration of contact.

The tests take into consideration and effect of device or its materials on the body or tissues and effect of body or tissues on the device.

While it is possible to correlate local reactions elicited by implant materials in animals and humans, the same can not be said of systemic effects or correlation with in vitro findings.

Metal alloys and coatings used for implantation rarely cause biocompatibility problem.

Polymers which are widely used in devices themselves are usually biocompatible but the additives, residual monomer contaminants may cause local or systemic reaction.

Biological tests usually performed on starting materials and some times on components are :

  • Sensitizing assay

  • Irritation tests

  • Cytotoxicity

  • Acute systemic toxicity

  • Haematocompatibility

  • Pyrogenicity

  • Implantation tests

  • Mutagenicity

  • Subchronic toxicity

  • Chronic toxicity

  • Carcinogenesis bioassay

  • Pharmacokinetics

Selection of biological tests depends on the nature and duration of contact. Additionally a device which comes into contact with broken skin or inserted into body cavity or implanted in the body needs to be sterile.

Back | Back To Top